EVALUATION METHOD

Understanding ANSI/CTA-2051-A Evaluations

Audiorama measures the electroacoustic performance of listening devices. For each measurement rated, a score of 5 out of 5 means that the applicable threshold specified in ANSI/CTA-2051-A is met. When the measurement deviates from this threshold in a negative direction, the score gradually decreases to 1.

How the score is calculated

The standard specifies the thresholds. Audiorama defines, for each measurement, an unfavorable deviation D that results in a score of 1 out of 5. Between the threshold and this deviation, an S-curve determines the score. A measurement halfway between them results in a score of 3 out of 5.

Example. The latency threshold is 15 ms and the D deviation is 35 ms. A latency of 15 ms is worth 5, a latency of 32.5 ms is worth 3, and a latency of 50 ms or more is worth 1. See the formula and rounding rules.

We calculate the unfavorable deviation e, then we bring the ratio back e/D between 0 and 1. This ratio is called bounded r.

Note = 3 + 2 × tanh[3 × (0.5 − r)] / tanh(1.5)

For an upper bound, e is the measurement minus the threshold. For cutoff frequencies, the deviation is calculated in octaves. It is equal to log2(low-cut filter / 250 Hz) or log2(high-cut threshold / high-cut), depending on the measurement.

The proposed calculation rounds scores down to one decimal place. When a threshold is exceeded, it caps the displayed score at 4.9 to reserve 5.0 for measures that meet the threshold.

Measures, Thresholds, and Scale

The table shows what each test measures, the threshold that yields a score of 5, and the values that yield scores of 3 or 1 using the spreadsheet's default settings. The frequencies corresponding to a score of 3 are rounded to the nearest hertz.

MesureWhat the test measures5 out of 53 out of 51 out of 5 starting at
Low-side cut-off (§5.1)How far does the system's response extend into the low frequencies?≤ 250 Hz354 Hz500 Hz and higher
High-voltage interruption (§5.1)How high does the frequency response extend, depending on the claimed class?Standard Band ≥ 5 kHz
; Wide Band ≥ 10 kHz
3,536 kHz
7,071 kHz
2.5 kHz and below
, 5 kHz and below
Frequency Regularity (§5.2)The most pronounced peak relative to the bands located two-thirds of an octave on either side.≤ 12 dB15 dB18 dB or more
Maximum Output of the OSPL90 (§5.3)The maximum output level measured at a 2 cm³ coupler for an input of 90 dB SPL.≤ 120 dB SPL123 dB SPL126 dB SPL and higher
Output THD+N, low level (§5.4.1)Total harmonic distortion and noise at 70 dB SPL at the output with a sine wave or at 67 dB SPL with pulsed noise.≤ 5 %17,5 %30% or more
Output THD+N, high level (§5.4.1)The same measurement at 100 dB SPL at the output with a sine wave or at 97 dB SPL with pulsed noise.≤ 5 %17,5 %30% or more
THD+N with high input (§5.4.2)Distortion and noise at an input of 100 dB SPL for a sine wave or 97 dB SPL for pulsed noise. The target output is 80 dB SPL.≤ 5 %17,5 %30% or more
Intrinsic Noise (§5.5)The noise produced by the device, expressed as the equivalent sound level at the input.≤ 32 dBA36 dBA40 dBA and above
Latency (§5.8)The time interval between the arrival of the sound at the input and its reproduction at the output.≤ 15 ms32.5 ms50 ms or more

The three THD+N tests use a 500-Hz test signal. Since no band class is specified, the high-pass cutoff is shown for descriptive purposes only, and only the low-pass cutoff is included in the bandwidth rating.

Low cut12345250354500Cutoff Frequency (Hz) High Cut — Standard Band1234553,5362,5Cutoff Frequency (kHz) High-pass filter — Wide Band12345107,0715Cutoff Frequency (kHz) Frequency Regularity12345121518Largest peak (dB)

Audiorama rating curves for bandwidth and frequency flatness. The dots represent ratings of 5, 3, and 1. The dotted lines indicate the applicable ANSI/CTA-2051-A thresholds. Crossover frequency deviations are measured in octaves; therefore, the midpoint is a geometric mean in frequency. The curves show the ratings before the spreadsheet was rounded.

Maximum Output OSPL9012345120123126Maximum Output (dB SPL) THD+N at the output, low level12345517,530THD+N (%) THD+N at the output, high level12345517,530THD+N (%) THD+N with high input12345517,530THD+N (%) Bruit propre12345323640Noise level at the input (dBA) Latence123451532,550Input-to-output delay (ms)

Audiorama rating curves for maximum output, distortion, self-noise, and latency. The dots represent ratings of 5, 3, and 1. The dotted lines indicate the applicable ANSI/CTA-2051-A thresholds. The three THD+N tests use the same curve under different measurement conditions. The curves show the scores before rounding in the spreadsheet.

How to Interpret the Overall Score

The overall score is the equally weighted average of six categories. For bandwidth and distortion, the lowest score among the required tests is used. This prevents a good result in one test from masking a failure in another test within the same category.

  • Bandwidth — minimum applicable cutoff frequencies.
  • Frequency regularity — magnitude of the most pronounced peak.
  • Maximum output — OSPL90 rating.
  • Distortion — the lowest of the three THD+N values.
  • Noise level — equivalent noise level at the input.
  • Latency — a measure of the delay.

The overall score is displayed when all six categories can be scored. If a required test is missing, the overall score field remains blank. Any instance of exceeding the limit is flagged, even if other tests remain to be completed or if the overall score is high.

Rating: 5 out of 5. The overall rating is based on the evaluated electroacoustic thresholds. The reported information and items to be verified at the point of sale are presented separately in the fact sheet.

What's Included in the Detailed Information Sheet

The data sheet lists the declared values, features, and information available at the point of sale. For a given feature, the data sheet may indicate that it is present, absent, not applicable, or undocumented.

High-Frequency Gain (§5.6)Average maximum gain at 1.0, 1.6, and 2.5 kHz for a 50 dB SPL input.
Battery Life and Battery (§5.7)Estimated runtime per cycle with optional features enabled, then disabled. Recommended battery type, if replaceable.
Radio Frequency Immunity (§5.9)Class M for microphone mode and Class T if the device has a telephone coil. The rating indicates whether the recommended M2 or T2 benchmark is met for the applicable mode.
Equalization (§5.10)Manual tone adjustments and automatic level-dependent processing.
Level-Dependent Gain (§5.11)Linear operation, single-band or multi-band compression, and the presence of wide-dynamic-range compression or limiting.
Improving the Signal-to-Noise Ratio (§5.12)Types of microphones listed, including omnidirectional, directional, and remote microphones.
Noise Reduction (§5.13)Availability and description of the feature.
Feedback Control (§5.14)Availability and description of the acoustic feedback hiss reduction feature.
Customization (§5.15)Type of adjustment and method used, such as selecting presets or adjusting based on the user's hearing characteristics.
Ear Coupling (§5.16)Open, closed, or adjustable fit. The product description may specify the tip or configuration.
Wireless Connectivity (§5.17)Supported modes, such as Bluetooth, Wi-Fi, or a telephone coil.

Information at the Point of Sale (§6)

Published bandwidth and maximum output, as well as the address where the device's measurements and specifications can be viewed.

The data sheet may also record an unfeasible test condition, the mode tested, the coupler used, or a deviation from the standard method.

Understanding Acoustic Transparency and Passive Attenuation Assessments

Acoustic Transparency and Passive Attenuation

In addition to the ANSI/CTA-2051-A measurements, Audiorama documents five experimental metrics proposed by Pinsonnault-Skvarenina et al. (2025). They compare sounds measured on an artificial head without the device, with the device activated in transparent mode, and with the device deactivated. These metrics are not ANSI/CTA-2051-A compliance requirements.

Good transparency allows you to hear your surroundings with minimal changes in sound level, frequency response, and directional characteristics. Passive attenuation, on the other hand, refers to the reduction in ambient sound when the device is turned off.

IndicatorWhat Is Being AssessedCalculation and ReferenceInterpretation
TI — Transparency IndexAverage difference in sound level in transparent modeTI = 1 − min(1, |T̄| / 10 dB), where T̄ is the average transparency across all directions and third-octave bands from 31.5 Hz to 20 kHz.1: average level unchanged; 0: average absolute deviation of at least 10 dB.
LocI — Localization IndexVariability in Transparency Depending on DirectionLocI = 1 − min(1, σ̄ / 6 dB), where σ̄ is the frequency-averaged standard deviation of the deviations between directions.1: Curves are identical regardless of direction; 0: Average variability of at least 6 dB. This is an indirect indicator, not a perceptual localization test.
FI — Flatness IndexSpectral Uniformity of Passive AttenuationFI = 1 − min(1,|pA|/ 3.5 dB/octave), wherepA is the regression slope of the passive attenuation as a function oflog2(f).1: zero slope; 0: absolute slope of at least 3.5 dB/octave. A zero slope does not guarantee the absence of local irregularities.
AI — Attenuation IndexOverall reduction in A-weighted sound levelAI = min(1, max(0, A / 25 dB)), where A =LA −L′A, for the defined test spectrum.1: Attenuation of at least 25 dB; a high value improves sound insulation, but not necessarily the perception of the environment.
fcI — Cutoff Frequency IndexExtension of transparency to high frequenciesDeterminefc, the first frequency above 1 kHz at which the average transparency differs by more than 10 dB from its average value between 31.5 Hz and 1 kHz; normalization in 20-kHz increments.A higher cutoff frequency is preferable. If no overshoot is detected below 20 kHz, record “≥ 20 kHz.”

Converting indicators to scores on a scale of 1 to 5

To ensure that the results are easy to read and consistent with other Audiorama results, each indicator is converted into a continuous score ranging from 1 (low performance) to 5 (high performance according to the criterion). The nonlinear conversion shown below is an Audiorama scoring convention, not a requirement of the ANSI/CTA-2051-A standard nor a scale validated in the original study.

N(q) = 3 + 2 × tanh[3 × (q − 0.5)] / tanh(1.5)

The variable q, which ranges from 0 to 1, represents the normalized performance for each indicator. This yields N(0) = 1, N(0.5) = 3, and N(1) = 5. Scores can be displayed with one decimal place.

NoteNormalized performance qRating of 5/5Note on Interpretation
Transparency (TI)q = 1 − min(1, |T̄| / 10 dB)Average transparency deviation of zeroA zero average can mask variations depending on the frequency.
Directional Consistency (LocI)q = 1 − min(1, σ̄ / 6 dB)Zero directional variabilityThis is not a psychoacoustic measure of localization.
Flatness of Attenuation (FI)q = 1 − min(1,|pA|/ 3.5 dB/oct)Zero regression slopeDoes not describe local irregularities in the spectrum.
Passive Insulation (AI)q = max(0, min(1, A / 25 dB))A-Weighted attenuation  ≥ 25 dB5/5 means “high insulation,” not “best for all uses.”
Frequency extension (fcI)q = max(0, min(1, (fc − 1 kHz) / 19 kHz))Cutoff frequency ≥ 20 kHzAudiorama normalization between 1 and 20 kHz, which differs from the wording in the article.

Important: The five scores reflect distinct characteristics. They are not automatically added together and should not be interpreted as a certification or as a direct measure of hearing benefit.

IT — Transparency12345−10−50510Average Transparency Deviation (dB) LocI — Directional Consistency12345036Average Directional Standard Deviation (dB) FI — Flatness of attenuation12345−3,503,5Roll-off slope (dB/oct) AI — Passive insulation12345012,525A-weighted attenuation (dB) fcI — Frequency extension12345110,520Cutoff frequency fc (kHz)

Protocol and Limitations on Interpretation

For each frequency f, we measure T(f) = 20log10|H_(on)(f)/H_(off)(f)| and Att(f) = 20log10|H_(off)(f)/H_(off)(f)|. The attenuation values are therefore generally negative under this transfer convention; the A-weighted level reduction is positive. The measurements depend on the coupling to the ear, the directions of incidence, the operating mode, and the test spectrum.

The TI, LocI, FI, AI, and fcI indices are research indicators, and their reference thresholds are not regulatory thresholds. The choice of an appropriate index depends on the intended use: natural listening, spatial perception, communication, or sound insulation. The Audiorama ratings from 1 to 5 listed above are suggested conversions of the research indices. They are distinct from the electroacoustic compliance rating.

References:

A. Pinsonnault-Skvarenina, H. Nélisse, F. Bonnet, M. Hotton, and J. Voix, “Evaluating the Performance of Direct-to-Consumer Hearing Devices: A Comparative Study of Electroacoustic, Acoustic Transparency, and Passive Attenuation Characteristics,” J. Acoust. Soc. Am., 158(5), 3781–3792 (2025), DOI: 10.1121/10.0039866.

ANSI/CTA-2051-A, Wearable Sound Amplifier Performance Criteria, June 2022, Sections 5 and 6. The curve, the D values, and the weighting of the overall score are based on the scoring system established by the Audiorama team.