The Principle
Insertion loss (IL) compares the sound level measured in the ear without the device (open ear) to that measured with the device inserted (occluded ear). The difference, expressed in dB, indicates how much sound the device blocks in each octave band.
The measurement method complies with the ANSI/ASA S12.42 standard.
The Test Bench
Diffuse field
Measurements are taken in the double-walled audiometric booth at the CRITIAS laboratory at ÉTS. Four full-range speakers generate uncorrelated pink noise. These signals create a diffuse field in the center of the booth. A reference microphone is placed at a fixed position in the booth, directly above the artificial head.
The stimulus is delivered at four levels: 75, 85, 95, and 105 dB SPL.
Artificial head
The device is inserted into the ears of an artificial head (Head and Torso Simulator, HATS). The simulator replicates the head, the outer ear, and the ear canal, with a microphone in place of each eardrum.
Equipment
- Type 5128-C artificial head (Brüel & Kjær, Nærum, Denmark)
- Type 4190 Reference Microphone (Brüel & Kjær, Nærum, Denmark)
- 4-Channel Microphone Power Supply, Model 2829 (Brüel & Kjær, Nærum, Denmark)
- 4 RB-81 full-range speakers (Klipsch, Indianapolis, IN, United States)
- HTR-6130 Power Amplifier (Yamaha, Hamamatsu, Japan)
- ICUSBAUDIO7D USB Audio Adapter (StarTech.com, London, ON, Canada)
- PXI-1033 Data Acquisition System (National Instruments, Austin, TX, United States)
- Synchronous multichannel data acquisition software (Multi-Channel Live Data Acquisition), developed using MATLAB (MathWorks, Natick, MA, United States)
Assembly of the Test Bench
Measurement Conditions
The sound signal is measured at the ear microphone in three device states:
An Open Ear
Without a device
Ear covered / uncovered
OTC inserted, turned off.
Blocked ear / on
OTC inserted, powered on (Active Noise Cancellation (ANC) enabled).
Each measurement is performed twice, with the device reinserted between measurements, as required by the ANSI/ASA S12.42 standard, in order to account for variability in device insertion.
Calculated Values
PIL
Passive Insertion Loss
It measures the difference between the open ear and the occluded ear / with the device turned off.
TIL
Total Insertion Loss
It is placed between the open ear and the occluded ear / with the device turned on (“on”).
GARLIC
Active Insertion Loss
Active mode contribution: TIL − PIL
Calculation of Attenuation
For each condition and each ear, the transfer function H(f) between the artificial head microphone and the reference microphone is calculated. The insertion loss is the ratio, in dB, of the transfer functions with and without the device:
IL(f) =20·log10 | H(f) (open ear ) / H(f) (occluded ear ) |
The results are expressed in octave bands ranging from 125 Hz to 8 kHz and are linearly averaged in dB across the two measurements and both ears.
References
- ANSI/ASA S12.42-2010 (R2020), Methods for the Measurement of Insertion Loss of Hearing Protection Devices in Continuous or Impulsive Noise Using Microphone-in-Real-Ear or Acoustic Test Fixture Procedures, Acoustical Society of America, Melville, NY, United States.